Corneal Topography and Tomography



Corneal Topography and Tomography


Jagadesh C. Reddy

Viral Juthani

Priscilla G. Fowler

Christopher J. Rapuano



PRINCIPLES, TECHNOLOGIES, AND CLINICAL APPLICATIONS OF CORNEAL TOPOGRAPHY/TOMOGRAPHY

More than two-thirds of the refractive power of the eye is derived from the air-tear film interface on the anterior corneal surface. Consequently, the cornea accounts for approximately 43 diopters (D) to the total 60D of refractive power of the eye. Both the anterior (90%) and the posterior (10%) surfaces contribute to the total power of cornea. The central cornea is steeper than the peripheral cornea (prolate) which has the advantage of reducing spherical aberrations. Any minute change in the shape of the cornea can cause a change in refractive power of the eye. The importance of the cornea on visual function has been the impetus for the development of instruments to measure the complex shape, thickness, and refractive power of the cornea. Corneal topography/tomography demonstrates the variations on the surface of the cornea. It is used to diagnose and follow pathology, plan refractive surgery, and evaluate postoperative results. Classic corneal topography is based on the principle of reflection of concentric lighted rings on the cornea. Over time, various corneal imaging instruments have been designed based on necessity and available technology. These include conventional keratometry, Placido disk imaging, raster photogrammetry elevation based technology, slitscanning topography, Scheimpflug imaging, ultrasound digital topography and optical coherence tomography (OCT). Raster photogrammetry, slit-scanning topography, Scheimpflug imaging, optical coherence tomography (OCT), and ultrasound digital topography are termed “tomography” systems due to creation of 3-dimensional image from a 2-dimensional image.

The keratometer and Placido disk are reflection-based methods. An object (multiple points or mires) of known diameter and distance from the cornea is projected onto the central corneal zone. The cornea acts like a convex mirror, reflecting light, and a virtual upright image is produced. The ratio of the image and object diameters is used to estimate the radius of curvature along a particular meridian (Fig. 112.1). Keratometers are highly accurate for regular spherocylindrical surfaces such as the central zone of a normal cornea, but they have limited value in irregular corneas. They measure the central 3 mm of the cornea, which only accounts for 6% of the entire surface. The keratometer provides no information as to the shape of the cornea either inside or outside the contour of the mires. In addition, standard keratometers have less accuracy for corneal powers below 36 D and above 50 D, although they have a theoretical range of 30 to 60 D.

A Placido disk reflects a series of concentric rings, or mires, off the cornea. Each mire produces an upright virtual image in the anterior chamber (about 4 mm behind the corneal vertex) of the subject’s eye.1 The distance between each mire to the corneal apex is compared to the image’s distance to the corneal apex. The ratio of image distance to mire distance determines the anterior corneal curvature along any particular meridian. Placido devices include photo-keratoscopes (Corneoscope) and computerized videokeratoscopes (e.g., TMS-Tomey Corporation, Nagoya, Japan; Optikon 2000-Rome, Italy; Magellan Mapper-Nidek, Fremont, California; Topolyzer-Alcon, Fort Worth, Texas; EyeSys 3000-EyeSys Vision Inc., Houston, Texas; ATLAS 9000-Carl Zeiss Meditec, Jena, Germany). Placido disk systems have distinct advantages over the keratometer. The photokeratoscope can measure 12 mires over 70% of the corneal surface and over an almost infinite diopteric range. The computerized videokeratoscope assumes that the corneal surface is spherical and of uniform refractive index; it can measure 15 to 38 mires over 95% of the corneal surface and has a theoretical range of 8 to 110 D.2

Videokeratoscopes display topographical information using a variety of maps. A picture of the cornea can also be provided as a raw image. This allows the clinician to recognize distortions of the tear film and poor image centration that may compromise the accuracy of the videokeratoscopy. The data are then presented as a color map with either an absolute or a normalized scale. On color maps, the warmer colors (red, orange, and yellow) represent the steeper areas; the cooler colors (blue and green) represent the flatter regions. The absolute scale is a fixed system in which each color is consistently assigned to a particular curvature or power. This facilitates the comparison of topographies between patients and over time in the same patient. Each videokeratoscope has its own absolute color map with fixed dioptric intervals covering the range of corneal powers. The normalized (relative) scale maximizes the utilization of all possible colors to evaluate a specific cornea. The color map with a normalized scale has narrower intervals between the assigned colors. Although the normalized scale allows the clinician to evaluate a particular topography with greater detail, it makes comparison between examinations difficult because the scale may have changed in range and dioptric interval. Another option available to the clinician is the adjustable scale, which allows the clinician to customize the step interval and dioptric range of the contours.






FIG. 112.1 Image formation by a convex mirror. Keratometry and keratoscopy utilize the property of the anterior corneal surface to reflect light, forming a virtual erect image within the anterior chamber. O, object; I, image; F, focal point; C, center of curvature of cornea; u, distance of object from cornea; v, distance of image from cornea; r, radius of curvature of cornea. (From Corbett MC, Rosen ES, O’Brart DPS. Corneal Topography: Principles and Applications. London: BMJ Books; 1999.)

Placido disk devices use axial (sagittal, global, average, or standard), tangential (instantaneous, true, or local), or refractive power maps to calculate radius of curvature (Fig. 112.2). The axial map measures the perpendicular distance from the tangent at a point to the optical axis and therefore has a spherical bias. This gives a global description of shape, but may be less accurate in the corneal periphery and in corneas with irregular surfaces. On the other hand, a tangential map displays the radius of curvature of a point with respect to its neighboring points along a specific meridian. It has less spherical bias than an axial map and therefore may be preferred to assess the corneal periphery or in irregular corneas. The refractive map uses Snell’s law to measure the focal power of the cornea.






FIG. 112.2 Global and local radius of curvature (ROC) and power. For the scale, radius of curvature or power can be selected. One is converted to the other using the standard keratometric index. A: Global/axial/sagittal measurements are made relative to the visual axis and therefore have a spherical bias. B: The same map expressed in terms of local/instantaneous/tangential measurements is more accurate over irregularities and in the corneal periphery. (From Corbett MC, Rosen ES, O’Brart DPS. Corneal Topography: Principles and Applications. London: BMJ Books; 1999.)

The videokeratoscope computer program generally gives the clinician several options for delineating the major and minor axes for a given cornea. The orthogonal axes mark the major and minor axes as measured at the central 3-mm zone. The zonal axes mark the major and minor axes at the 3-mm as well as the 5- and 7-mm zones. This allows the clinician to quickly assess whether there is a shift in the major and minor axes with progression toward the corneal periphery. The instantaneous axes option draws a continuous line from the center of the cornea to the periphery, showing the true major and minor axes at any given diameter (Fig. 112.3).

Data can also be manipulated by videokeratoscope computer programs to glean desired information. Crosssectional maps display the theoretical elevation and depression of the anterior corneal surface along any given meridian. Difference maps can compare pre- and posttreatment corneal topographies by subtracting the first map from the second. Statistical indices, which are often used in clinical studies, can also be obtained using computer programs.

Raster photogrammetry (e.g., PAR-PAR Technology, New Hartford, New York) is another projection-based system that produces an elevation map of the cornea. Its advantages are that it measures the entire corneal surface, up to 12 mm, can also measure irregular or deepithelialized or keratectomized corneas, and can be used intraoperatively. A projector casts a grid onto a corneal surface coated with fluorescein dye. The fluorescein dye produces light emissions from the surface of the cornea. Two or more camera systems at different angles to the cornea from the projector image the grid (fluorescence-based photography). The rays intersect in 3-dimensional (3D) space, creating the elevation map. It has the added advantage of rapid image acquisition and being less sensitive to focusing.3, 4 and 5






FIG. 112.3 Orthogonal axes mark the major and minor axes as measured at the central 3-mm zone. Zonal axes mark major and minor axes at the 3-, 5-, and 7-mm zones. The instantaneous axes are a continuous line showing the true major axes at any given diameter.

Scanning slit topography (e.g., Orbscan IIz; Bausch & Lomb, Rochester, New York) is a projection-based method that uses a series of slit-beam images to generate data regarding anterior surface curvature, posterior surface curvature, and pachymetry. Therefore, the cornea is represented as a 3D structure. Two technologies are used in this particular system, the scanning slit-beam system and the Placido disc system. A slit lamp projects a beam at 45 degrees onto the cornea. Twenty slits are projected sequentially on the eye from the left side, and 20 slits from the right, for a total of 40 slits that produce 240 data points per slit (Fig. 112.4). These points form the source for the mathematical calculation and are used to compute four elements of the anterior segment: anterior and posterior elevations, corneal power, and pachymetry maps. The elevations and depressions are recorded in 5 µm increments. Height maps have a colored reference scale and display elevations in warm colors and depressions in cool colors. The distance between the anterior and the posterior surfaces is the corneal thickness. Corneal thickness measured by the Orbscan is 7% thicker than the gold standard ultrasound pachymetry measurements6, 7 (Fig. 112.5).






FIG. 112.4 A and B: Slit scans projecting onto cornea from left and right sides of video axis. C: Overlapping slit scans in the central 5-mm zone of the cornea.

The Orbscan is also capable of calculating corneal power. Mean power maps show the variation of spherical power by calculating the arithmetic average of the principal curvatures for each data point. Astigmatic power maps represent absolute local cylinder over the corneal surface. It is obtained by calculating the arithmetic difference of the principal curvatures for each data point. In addition, axial, tangential, and Snell power maps are available, based on the same principles as the Placido disk systems.






FIG. 112.5 The principle behind the slit topography projection-based method involves ray-trace triangulation. (Orbscan, ORBTEK, Inc., Salt Lake City, Utah.)

Scheimpflug imaging attains a wide depth of focus, providing sharp images that include information from the anterior corneal surface through to the posterior crystalline lens capsule (anterior segment of the eye). There are two Scheimpflug systems available on the market, the Pentacam (Oculus Inc., Wetzlar, Germany) and Galilei dual Scheimpflug analyzer (Ziemer Group, Port, Switzerland).

The Pentacam consists of two cameras, a central camera monitoring the fixation and the Scheimpflug camera scanning the optical cross-sections of the anterior segment.8 In a span of 2 seconds, 25 to 50 images are captured by the camera yielding about 12,500 to 25,000 data points. These data points are used in computing anterior and posterior sagittal and tangential curvature, anterior and posterior elevation maps, corneal pachymetry from limbus-limbus, anterior chamber depth, anterior chamber angles, corneal and anterior chamber volumes, pupil diameter, and lens thickness. In addition, the equivalent keratometry reading calculated by Pentacam can be used for the corneal power after kerato-refractive surgery to calculate intraocular lens power.

The Galilei dual Scheimpflug analyzer integrates a Placido disk and a dual rotating Scheimpflug system (two cameras 180 degrees apart) to compensate for error associated with single scans at an oblique angle. During the rotating scan, Placido disk and Scheimpflug images are simultaneously acquired to obtain the information on the curvature and elevation of the cornea, respectively. In a span of 1 to 2 seconds approximately 122,000 data points are measured per scan. The dual camera configuration captures two Scheimpflug slit images from opposite sides of the slit beam and simultaneously tracks decentration due to eye movements, resulting in highly reliable pachymetry and posterior curvature data. The quality of the acquired scans need to satisfy the minimum percentage required by the instrument; the percentages are automatically calculated by the internal software for different parameters, such as motion compensation, Placido, Scheimpflug images, and motion distance (Fig. 112.6).






FIG. 112.6 Quality analysis report on the Galilei dual Scheimpflug imaging unit depicting the actual (measured) and recommended quality percentages. Based on these percentages there is possibility of either discarding or accepting the measurement.






FIG. 112.7 Four-view refractive display on the Galilei.

The refractive report consists of four fixed maps (anterior axial curvature, pachymetry map, anterior elevationBest Fit Sphere, posterior elevation-Best Fit Sphere) and a set of parameters that provide a comprehensive overview of the cornea (Fig. 112.7). The various parameters include:



  • Anterior Curvature: SimKavg is the arithmetic mean or average of the steep and flat axis (SimKf, SimKs), which are calculated from the pair of meridians 90 degrees apart with the greatest difference in average power, from 0.5 to 2.0 mm distance from the center.


  • Posterior curvature parameters are calculated with the indices of refraction of the cornea (1.376) and aqueous humor (1.336).


  • Total Corneal Refractive Power (mean, flat, steep, astigmatism, central Avg [0 to 4 mm], Paracentral Avg [4 to 7 mm], Peripheral Avg [7 to 10 mm]): It is calculated by ray-tracing through the anterior surface and posterior surface using Snell’s Law and the thin lens equations.


  • Corneal pachymetry: The mean corneal pachymetry at the central zone (0.0 to 4.0 mm), paracentral zone (4.0 to 7.0 mm), peripheral zone (7.0 to 8.0 mm), and thinnest pachymetry.


  • Limbus (Nasal-Temporal, Superior-Inferior): The maximum length in the horizontal and vertical directions of the ellipse respectively.

The keratoconus report consists of keratoconus indices (Table 112.1), which are based on formulas of Placido device, and simulated on the Galilei by translating the ring numbers used in the original Placido equations into corneal regions in the Galilei9, 10 (Fig. 112.8).

Wavefront maps of the total cornea (front and back surfaces) are displayed in micron as well as in diopter. The individual coefficients are displayed for the second, third, and fourth order aberrations; the orders five to eight are displayed in a summarized fashion (Fig. 112.9).

The IOL power report shows the anterior chamber depth when the pupil is well dilated, volume and iridocorneal angle (temporal, superior, nasal, and inferior quadrants). These values are measured between the corneal endothelium and the anterior surface of the lens; the volume is calculated over a diameter of 8.0 mm (Fig. 112.10).

Difference maps (A – B difference) can be used to compare various parameters (axial curvature, anterior instantaneous curvature, refractive power, anterior height difference, corneal pachymetry, posterior axial curvature, posterior instantaneous curvature, posterior height difference, and total corneal power) between two visits to follow a specific pathology or to compare pre- and postoperative results (Fig. 112.11).

Galilei also has the facility to connect to FEMTO LDV laser using Z Lasik report. Scanning the report in by the FEMTO LDV helps in transferring the patient’s data from Galilei to FEMTO LDV electronically, thus eliminating sources of error during data entry and also minimizing the laser setup time.








TABLE 112-1 KCN indices measured by the Galilei






Inferior-superior index (I-S)


Difference between the inferior and superior average dioptric values approximately 3 mm peripheral to the corneal vertex (center of the map)


Surface asymmetry index (SAI)


Centrally weighted average of the summation of differences in corneal power between corresponding points 180° apart on all meridians


Surface regularity index (SRI) Difference in power gradient between successive points on a hemimeridian that is assigned a positive value and added to the running sum. This sum is divided by the number of points that went into the sum


Irregular astigmatism index (IAI)


Average summation of area-corrected dioptric variations along every semi-meridian for the entire analyzed surface and normalized by the average corneal power and number of all measured points


Area analyzed (AA)


Ratio of the actual data area to the area circumscribed by a circle 4.5 mm in radius


Differential sector index (DSI)


Greatest difference in average area-corrected power between any two 45° sectors


Opposite sector index (OSI)


Greatest difference in average area-corrected power between opposite 45° sectors


Central/surround index (CSI)


The simulated center/surround index reports the difference between the average area corrected power between the central area (3-mm diameter) and an annulus surrounding the central area (3-6 mm)


Average central power (ACP)


ACP is simulated as the average dioptric power of all points within the central 3 mm


Standard deviation powers (SDP)


SDP is simulated as the standard deviation of all measured powers present on the map


Keratoconus prediction index (KPI)


The keratoconus prediction index is a compilation index of DSI, OSI, CSI, SAI, SimK1, SimK2, IAI, and AA. It is simulated on the Galilei as a percent probability based on a 100-point scale







FIG. 112.8 Four-view keratoconus display on the Galilei showing cone localization on the elevation and pachymetry maps.

Wang et al.11 have shown that the Galilei had excellent repeatability for corneal powers, pachymetry, and anterior chamber depth measurements, and Menassa et al.12 have shown low intraobserver and interobserver variation for central corneal thickness and power. The central corneal thickness measurements made with the Galilei showed good correlation and agreement with those made with ultrasound.13

The Artemis very high-frequency (VHF) digital ultrasound arc-scanner (ArcScan Inc., Morrison, Colorado) was designed for imaging and measuring the whole anterior segment to improve anatomical diagnosis for surgical planning and postoperative diagnostic monitoring. Using a broadband 50 MHz VHF ultrasound transducer, Bscans are acquired as arcs that follow the surface contour of anterior segment structures. Digitized ultrasound data are then transformed using digital signal processing technology (I-scan). Thickness profiles are calculated based on data from four meridional B-scans, comprised of eight semi-meridians. Performing a 3-dimensional scan set with the Artemis 1 takes approximately 2 to 3 minutes per eye. The axial measurement precision for the entire anterior segment (15-mm-wide) is approximately 1 µm, while the lateral precision is 0.12 mm for measuring angle-toangle dimensions, and 0.22 mm for sulcus-to-sulcus dimensions. The resolution of the Artemis, when set to scan the cornea, is sufficient to distinguish individual corneal layers such as the epithelium, stromal component of a LASIK flap, residual stromal bed. Work done by Reinstein et al.14, 15 and 16 on the utility of the VHF digital ultrasound arcscanner to differentiate layers of the cornea and anterior segment highlights the usefulness of this technology in differentiating normal and keratoconic corneas, changes induced in the cornea as a consequence of LASIK, and also helps in anatomical surgical planning and postoperative monitoring of phakic IOLs.






FIG. 112.9 Galilei generated Zernike polynomials and wavefront maps of the entire cornea.

Visante anterior segment optical coherence tomography (Carl Zeiss Meditec, Inc., Dublin, California) imaging is based on the principle of measuring the delay of light (wavelength 1310 nm) reflected from tissue structures. Optical coherence tomography (OCT) employs low-coherence interferometry to compare the delay of tissue reflections against a reference reflection. The pachymetry scan can be either in standard pachmetry map (8 sequential images) or global pachymetry map (16 sequential images). The relative pachymetry map compares the patient’s pachymetry map with the standard normal cornea (550 µm) (Fig. 112.12). The pachymetry data (central 5 mm) can be used to help diagnose keratoconus based on the average thickness of the inferior octant minus the average thickness of the superior octant (cutoff: >-45 µm), the average thickness of the inferotemporal octant minus the average thickness of the superonasal octant (cutoff: >-45 µm), minimum pachymetry of <470 µm, minimum-maximum pachymetry >-100 µm. When Visante OCT (pachymetry, posterior surface, power) is linked with Humphrey ATLAS corneal topographer (Placido disk-axial, instantaneous, anterior elevation, wavefront), all the parameters of the cornea can be analyzed.17, 18

The latest technology is that of wavefront sensors, which evaluate the sum of refractive effects of the cornea, lens, media, and retina. Wavefront sensor technology dates from the 1970s, when Josef Bille, at the University of Heidelberg, applied astronomical principles to the study of the eye. Lightwaves from distant stars traveling through the atmosphere and entering a telescopic lens should be virtually flat. However, interference from the atmosphere produces aberrations in the wavefront. Bille recomposed the imperfect wavefronts into a perfectly round image through the use of different mirrors. Fritz Zernike (1934) described a set of polynomials representing a particular type of optical aberration. Aberrations can also be represented using Fourier series. Fourier full reconstruction is faster and more accurate than Zernike reconstruction, and better represents ocular aberrations of highly aberrated eyes.19 Today, wavefront technology is designed to evaluate refractive errors and correct optical aberrations, and it has been applied to laser refractive surgery. Wavefront technologies commercially available to measure aberrations include Tscherning aberroscope, differential skiascopy, Hartmann-Shack sensor, and ray tracing. In Hartmann-Shack technology, a laser beam is projected through the cornea, lens, and media onto the fovea; the reflected beam of light produces wavefronts that are captured by a grid of lenses and sensors that lie in a conjugate plane to the cornea. The wavefront’s angle of incidence to the lens is recorded by the sensor. The shape of the wave can be reconstructed from the direction in which the wave is traveling at each of the grid lenses (Fig. 112.13). Regions of the wave that are out of phase with a reference point such as the front of the wave can be detected. A computer program develops a color map representing total ametropia at different points. Wavefront maps can determine whether the wave emitted from one part of the optical system is out of phase with a wave emitted from another region. The discrepancy in phases can be measured in microns and corresponds to the depth of ablation necessary to produce a flat wavefront. A lens can be formulated to compensate for the aberration, creating a customized ablation pattern. In ray tracing technology, forward aberrations of light passing into the eye are measured, which are considered more physiologic in analyzing vision. Ray tracing technology has certain advantages over other technologies as laser spots projected through the entrance pupil can be controlled with the software; there is no confusion in the analysis between the entrance pupil and the reflected location from the retina due to rapid sequential capturing of the data.20, 21, 22, 23 and 24






FIG. 112.10 Corneal diameter, anterior chamber, and pupil indices measured by Galilei.






FIG. 112.11 Comparison of two examinations (pre- and post-epi-LASIK) using Galilei.






FIG. 112.12 Pachymetry mapping using Visante Optical Coherence Tomography.






FIG. 112.13 Wavefront technology evaluates the entire optical system. The eye is illuminated by a plane wave. In an ideal eye, the wave traveling inside this organ is a spherical wave that provides a bright point on the fovea. In reality, the wavefront is not spherical. If the fovea acts as a point source, it provides a divergent spherical wave that exits the eye as a plane wave, which in reality is a deformed wave. (From Haman H. A quick method for analyzing Hartmann-Shack patterns: application to refractive surgery. J Refract Surg. 2000;16:S636-S642.)


NORMAL CORNEAL TOPOGRAPHY

The cornea is not a perfect sphere. The central 4-mm zone of the normal cornea is fairly spherical, but the periphery is aspherical. The normal cornea is steepest centrally and becomes progressively flatter in the periphery. This configuration is referred to as prolate. The reverse pattern of central flatness and peripheral steepness is referred to as oblate. The oblate configuration is not seen in normal corneas.

The general population has a variety of corneal curvatures that are compatible with good vision and within the normal spectrum. Five videokeratoscopic patterns lie on the normal spectrum of corneal curvature: round (23%), oval (21%), symmetric bow tie (18%), asymmetric bow tie (32%) (in which the ends of the bow tie are of different sizes), and irregular (7%) (in which the ends of the bow tie are at an angle to one another) (Fig. 112.14).25 In addition, there is diurnal variation due to stromal edema during sleep, menstrual variation, and lifetime variation. Infants usually have spherical corneas. During childhood and adolescence, regular with-the-rule astigmatism develops in about 90% of individuals. This reverses with subsequent aging processes. Diurnal and menstrual variations are generally subclinical.26






FIG. 112.14 Five qualitative patterns of normal corneal topography using a normalized scale. Top left, round; top center, oval; top right, symmetric bow-tie; bottom left, asymmetric bow tie; bottom center, irregular. In the normalized scale (bottom right) the range of dioptric power represented by each color varies among eyes, depending on the degree of corneal asphericity. (Courtesy of Bogan SJ, Waring GO, Ibrahim O, et al. Classification of normal corneal topography based on computer-assisted videokeratography. Arch Ophthalmol. 1990;108:945-949.)

Corneal elevation patterns on the Orbscan are measured compared to a best-fit sphere. Using the Orbscan, Liu et al.27 classified anterior and posterior elevation patterns seen among normals into five categories. Anterior corneal elevation patterns occur with the following frequency: the island (72%), incomplete ridge (20%), regular ridge (4%), irregular ridge (2%), and unclassified (2%). Posterior elevation patterns occur with a different order of frequency: the island (33%), regular ridge (30%), incomplete ridge (24%), irregular ridge (13%), and unclassified (0%).

The anterior corneal surface has an average power of 49.50 D, and the posterior corneal surface has an average power of -6.00 D, rendering a total corneal power of approximately 43.50 D. Mean keratometric astigmatism is 0.90 D. Central corneal thickness is 560 µm and peripheral corneal thickness is 1,200 µm. Pachymetry patterns were classified by Liu et al.27 into four categories: round (41%), oval (48%), decentered round (2%), and decentered oval (9%).


TOPOGRAPHY AND TOMOGRAPHY OF CORNEAL DISEASES

Corneal topography and tomography are often useful clinically for the detection of disease, evaluating severity, and assessing response to treatment. Many disorders of the cornea are identified by subtle changes in topography/tomography often before they cause visual symptoms or visible signs on slit-lamp biomicroscopy. The most classic corneal disorders identified by corneal topography/tomography are the corneal ectasias, including keratoconus, pellucid marginal degeneration, and post-LASIK ectasia. In addition, ocular surface masses, peripheral corneal thinning disorders, and corneal scarring can create changes in topography that deviate from normal patterns.


Stromal Ectasias

Corneal ectasias are noninflammatory disorders that are characterized by thinning and protrusion of the corneal stroma, resulting in changes in shape. This family of diseases consists of keratoconus, keratoglobus, and pellucid marginal degeneration. Normally, the stromal collagen lamellae run circumferentially in the corneal periphery, producing a spherical shape. However, progressive thinning of the stroma leads to flattening of the corneal curvature along that meridian. This induces the peripheral ring of collagen lamellae to assume a more oval shape and transmits a compressive force to the lamellae that are 90 degrees away, resulting in corneal steepening in that meridian. This mechanism is known as biomechanical coupling. Furthermore, the normal intraocular pressure at the site of weakness causes protrusion of the cornea.


Keratoconus

Keratoconus is a bilateral but typically asymmetric corneal ectasia that has onset in the late teens and usually progresses slowly over many years. Patients typically have a history of progressive myopia, oblique astigmatism, and reduction of spectacle-corrected visual acuity. Prior to the introduction of methods to objectively assess corneal topography, the diagnosis was based on history and the presence of clinical signs. In mild disease, however, these clinical signs are subtle or altogether absent. The advent of refractive surgery has made the detection of subclinical keratoconus increasingly important in order to prevent the surgical treatment of these eyes. Keratoconus has an incidence of 1 in 2,000 in the general population, but is detected in 5% of myopes who present for refractive surgery evaluation.28

Keratoconus can be categorized according to the severity of power (mild, moderate, severe); location of cone (superior, central, inferior); and shape of cone (oval, globus, nipple) (Fig. 112.15). Corneal thinning most commonly occurs in the inferocentral cornea, and protrusion also occurs in this region. The point of maximal protrusion is referred to as the apex of the cone. The steepest corneal slope lies just peripheral to the apex (usually inferior in central cones). The region of smallest radius of curvature (therefore the greatest corneal power) lies between the cone’s apex and its steepest slope. The mechanism of biomechanical coupling causes the flattest meridian to be approximately horizontal and the steepest meridian to lie close to the vertical meridian. Placido disk-based videokeratographs mirror this distortion by producing mires that are typically oval. The distance between rings is smallest at the steepest corneal slope and farthest apart superiorly where the cornea is flattest. Tangential curvature maps of projection-based systems provide additional information. On these maps the steepest slope is easily located as being inferior to the apex, producing an asymmetric bow tie. This corresponds to the exaggerated prolate shape of the keratoconic eye. Projection-based systems can be used to locate the apex of the cone on elevation maps as the highest point. The apex is surrounded by concentric zones of decreasing elevation. A comparison of anterior and posterior elevation maps reveals that there is a greater change in height from periphery to central cornea posteriorly than anteriorly. On tangential curvature maps, the apex of a cone has a slope of zero. Protrusions such as keratoconus nodules (proud nebulae) also have a slope of zero.