Intraocular Pressure and Wound Status in Eyes Immediately After Scleral Tunnel Incision and Clear Corneal Incision Cataract Surgery




Purpose


To compare intraocular pressure (IOP) and wound architecture immediately after cataract surgery using a scleral tunnel incision or clear corneal incision.


Design


Prospective randomized clinical trial.


Methods


Bilateral eyes of 64 patients undergoing phacoemulsification were randomized to 2 groups: 2.4 mm scleral tunnel incision or 2.4 mm clear corneal incision. IOP was adjusted upon completion of surgery to between 15 and 30 mm Hg. The wound was hydrated in 87.5% of eyes in the clear corneal incision group. IOP was measured using a rebound tonometer preoperatively; at the conclusion of surgery; and at 5, 10, 15, 30, and 60 minutes postoperatively. The Seidel test and anterior segment optical coherence tomography (AS OCT) were performed at 30 minutes postoperatively.


Results


In the scleral tunnel incision group, mean IOP decreased to the preoperative level within 15 minutes postoperatively and did not change significantly for up to 60 minutes. In the clear corneal incision group, IOP decreased to lower than the preoperative IOP at 30 minutes postoperatively. Mean IOP was significantly higher in the scleral tunnel incision group than in the clear corneal incision group throughout the postoperative period ( P ≤ .0339). Hypotony (<10 mm Hg) was observed in 7 eyes (10.9%) that underwent clear corneal incision and in no eyes that underwent scleral tunnel incision ( P = .0131). Wound leakage and loss of wound coaptation were not observed in any eyes.


Conclusions


IOP was significantly higher immediately after scleral tunnel incision without hydration than after clear corneal incision with hydration, and both incisions closed within 30 minutes postoperatively without leakage based on AS OCT.


Endophthalmitis occurs more frequently after clear corneal incision cataract surgery than after scleral tunnel incision cataract surgery. Previous studies revealed that bacteria-sized particles flow into the anterior chamber through the clear corneal incision at a high incidence in the immediate or early postoperative period when intraocular pressure (IOP) is reduced. Further, recent studies using anterior segment optical coherence tomography (AS OCT) showed that a low IOP in the immediate postoperative period is associated with incomplete clear corneal incision wound closure. Thus, hypotony immediately or early after cataract surgery is a major risk factor for anterior chamber contamination.


Some studies evaluated IOP immediately after clear corneal incision cataract surgery. Shingleton and associates reported that hypotony occurs in approximately 20% of eyes at 30 minutes after clear corneal incision. Our previous study demonstrated that the IOP normalizes and stabilizes within 15 minutes after clear corneal incision, and clear corneal wounds close within 60 minutes based on AS OCT. No studies to date, however, have examined the temporal changes in IOP and wound architecture immediately after scleral tunnel incision cataract surgery. Corneal shape changes are smaller after scleral tunnel incision than after clear corneal incision when the incision length is greater than 2.5 mm. Furthermore, Ernest and associates reported, using cadaver eyes, that wound strength of a square scleral tunnel incision is stronger than that of a rectangular clear corneal incision. Accordingly, we assumed that the IOP fluctuation would be less and wound tightness more secure after scleral tunnel incision than after clear corneal incision.


The purpose of the present study was to compare the longitudinal changes in IOP and wound integrity between eyes after scleral tunnel incision and eyes after clear corneal incision. Adjustment of the IOP to normal range was achieved by corneal wound hydration and injection of irrigating solution in eyes that underwent clear corneal incision, whereas eyes that underwent scleral tunnel incision did not undergo corneoscleral wound hydration.


Methods


Patients


This study was a prospective randomized clinical trial. A clinical research coordinator began screening all consecutive patients scheduled for bilateral cataract surgery at the Hayashi Eye Hospital on May 10, 2013. Exclusion criteria were eyes with any pathology of the cornea, vitreous body, macula, or optic nerve; eyes with glaucoma or ocular hypertension; eyes with pseudoexfoliation syndrome; eyes scheduled for planned extracapsular cataract extraction; a history of previous ocular surgery or inflammation; patients with diabetes; patient refusal; and any anticipated difficulties with examination or follow-up. Eyes that were included in another study were excluded from the present study. Screening was continued until 64 patients were recruited on December 9, 2013. This research adhered to the tenets of the Declaration of Helsinki. The Institutional Review Board (IRB)/Ethics Committee of the Hayashi Eye Hospital, Fukuoka, Japan, at which the present study was conducted, approved the study protocol, and all patients provided written informed consent to participate in the present study.


Sample Size


The study sample size was determined to have a statistical power of more than 90% to detect a clinically meaningful magnitude of difference in IOP between eyes after scleral tunnel incision and eyes after clear corneal incision. Assuming that an IOP difference of 1 mm Hg was clinically meaningful, we calculated that 51 eyes per group were necessary based on standard deviation of previous study. Assuming a possible 20% loss to follow-up, we calculated that 64 eyes were required for each group.


Randomization


All enrolled patients were randomly assigned to 1 of 2 groups the day before surgery: (1) left eyes that were to undergo a 2.4 mm scleral tunnel incision group and right eyes that were to undergo a 2.4 mm clear corneal incision group, and (2) right eyes that were to undergo scleral tunnel incision and left eyes that were to undergo clear corneal incision. The clinical research coordinator for this trial generated a randomization code with equal numbers (1:1 ratio) using computer software, and assigned each patient to 1 of the 2 groups according to the computer-generated randomization code. The group to which the patients were assigned was directly communicated by the coordinator to a member of the operating room staff who prepared the intraocular lens (IOL). The surgeon was informed about the type of surgery just before surgery. Data were collected by the coordinator at the Hayashi Eye Hospital. To ensure allocation concealment, the coordinator kept the assignment schedule until all data were collected (December 12, 2013). All patients, examiners, and the data analyst were unaware of the group to which the patients had been assigned.


Surgical Techniques


All surgeries were performed by a single surgeon (K.H.) using the surgical procedures described previously. To equalize the effect of local anesthesia on IOP reduction between groups and to avoid eyelid pressure owing to patient tension, each eye received a 2 mL retrobulbar injection of 2% xylocaine prior to the preoperative measurement of IOP. For phacoemulsification surgery, the surgeon made a horizontal incision in eyes having against-the-rule or oblique corneal astigmatism, and a superior incision in eyes having with-the-rule astigmatism.


Scleral tunnel incision cataract surgery


For the scleral tunnel incision, 2 side ports were first created with a 0.6 mm diamond knife approximately 90 degrees away from the center of the main incision. Through 1 side port, a continuous curvilinear capsulorrhexis measuring about 5.5 mm in diameter was accomplished using a bent 25 gauge needle. After dissection of the conjunctiva and before incision, wet-field cauterization was gently applied at the bleeding sites of the superficial episcleral vessels. The surgeon then made a 2.5 mm straight incision that was 1.5 mm posterior to the limbus, and a scleral tunnel was dissected up to 1.0 mm anterior to the limbus with a diamond crescent knife. The anterior chamber entry was created with a 2.4 mm steel keratome. The intended length of the scleral tunnel incision was approximately 2.5 mm. After hydrodissection, phacoemulsification of the nucleus and aspiration of residual cortex was performed. The lens capsule was inflated with 1% sodium hyaluronate (Healon; Abbott Medical Optics, Santa Ana, California, USA), after which a single-piece hydrophobic acrylic IOL (SN60WF; Alcon Laboratories, Fort Worth, Texas, USA) was placed into the capsular bag using a Monarch II injector with a C cartridge (Alcon). At the conclusion of surgery, balanced saline solution was injected into the anterior chamber using a cannula through a side port, and the IOP was measured using a rebound tonometer (ICare; Tiolat, Helsinki, Finland). The side ports were hydrated at the conclusion of surgery. An examiner, trained in using the ICare tonometer, measured the IOP. Because the IOP was 15 mm Hg or higher in all eyes, wound hydration was not performed. Finally, the IOP was adjusted to range between 15 and 30 mm Hg by reinjection of balanced saline solution, or by removing anterior chamber fluid through a side port using a cannula.


Clear corneal incision cataract surgery


For the clear corneal incision, 2 side ports were first created with a 0.6 mm diamond knife at approximately 90 degrees away from the center of the main incision. Through a side port, a continuous curvilinear capsulorrhexis was accomplished using a bent needle. Then, a 2.4 mm single-plane clear corneal incision was created from the posterior margin of the cornea with a stainless steel keratome for phacoemulsification and implantation of the IOL. The intended length of the clear corneal incision was approximately 1.5 mm. At the conclusion of surgery, balanced saline solution was injected through a side port into the anterior chamber, and the IOP was measured. Eyes in which the IOP did not reach at least 15 mm Hg underwent stromal hydration around the wound. Finally, the IOP was adjusted to range between 15 and 30 mm Hg by reinjecting balanced saline solution, or by removing anterior chamber fluid through a side port using a cannula.


In this series, all surgeries were uneventful, and all IOLs were implanted in the capsular bag without sutures. The preoperative and postoperative medication and anesthesia were the same between groups, and examiners could not distinguish the type of incision used until the AS OCT examination was performed.


Outcome Measures


IOP of each eye was measured using the ICare rebound tonometer preoperatively; at the conclusion of surgery; and at 5, 10, 15, 30, and 60 minutes postoperatively. The details of this tonometer were described previously. Briefly, the ICare tonometer includes a solenoid magnetized probe and processing electronics. The probe moves toward the cornea at a speed of approximately 0.2 m per second. After the initial propulsion pulse is completed, the probe impact decelerates and rebounds from the corneal surface. Signal processing electronics and microcontrollers register the probe’s deceleration time on corneal impact. The software is preprogrammed for 6 measurements; the highest and lowest readings are automatically discarded, and mean IOP is calculated from the remaining readings. The software is able to detect if an incorrect measurement is obtained, such as the absence of probe movement, eye untouched, low probe speed, incorrect distance of the probe to the central cornea, or incorrect positioning of the probe. In these cases, the tonometer shows an error message and does not accept the readings as correct. When measuring a patient in the supine position, an arrow appears in the display that indicates the vertical position of the device. In addition, the ICare tonometer considers the relationship among all measurements obtained by estimating the standard deviation to ensure a coherent final result. When the device detects the existence of any discrepancy among measurements, an error sign is displayed. In the present study, IOP was measured with the patients lying in the supine position, without topical anesthesia. To measure the IOP accurately, the device was placed near the patient’s eye with the patient’s forehead used as a base support, and the tip of the probe was maintained at a distance of approximately 3–7 mm from the central cornea. The same examiner performed all IOP measurements for each patient. To ensure the reliability of the IOP readings, measurements were repeated 3 times, and the mean value was used for analysis. When any type of error sign was displayed or in the event of a discrepancy between 1 IOP reading and the other 2 readings, the reading was excluded and another measurement was performed. The reliability and reproducibility of the data obtained using the ICare tonometer were confirmed previously.


The wound architecture of the scleral tunnel incision and clear corneal incision in each eye was examined using AS OCT (Casia; Tomey, Nagoya, Japan) at approximately 30 minutes postoperatively. The AS OCT can scan across the anterior segment of the eye, traversing the main incisions and side ports, which show the scleral tunnel incision and clear corneal incision in profile. The AS OCT image of the incisions was exported for analysis. In the present study, 5 architectural wound features were used to objectively describe the status of the incisions according to the classification reported by Calladine and Packard : (1) wound gaping at the epithelial side, (2) wound gaping at the endothelial side, (3) misalignment of the roof and floor of the incision at the endothelial side, (4) local detachment of Descemet membrane, and (5) loss of coaptation along the stromal tunnel. In the present study, we defined wound opening as a loss of coaptation along the stromal tunnel.


Two physicians (T.T., M.Y.) performed the Seidel test at approximately 30 minutes postoperatively using slit-lamp biomicroscopy. Flare intensity was measured at 45 minutes postoperatively using a flare meter (FC-1000; Kowa, Tokyo, Japan). Central corneal thickness was examined preoperatively and at 30 minutes postoperatively using the AS OCT. The keratometric cylinder and refractive spherical as well as cylindrical powers were examined using an auto kerato-refractometer (KR-7100; Topcon, Tokyo, Japan) preoperatively and at 1 day postoperatively. Manifest spherical equivalent value was determined as the spherical power plus half the cylindrical power. Uncorrected and distance-corrected visual acuity were measured on decimal charts preoperatively and at 1 day postoperatively, and decimal visual acuity was converted to the logarithm of minimal angle of resolution (logMAR) scale for statistical analyses. Three experienced ophthalmic technicians who were unaware of the purpose of the study performed all examinations.


Statistical Analysis


The IOP, flare intensity, visual acuity, and other continuous variables between the scleral tunnel incision and clear corneal incision groups were compared using the unpaired t test or Mann-Whitney U test based on the results of the Shapiro-Wilk test. Discrete variables were compared using the χ 2 test or Fisher exact test. Longitudinal changes in IOP in each group were compared using a 1-way analysis of variance (ANOVA). When a statistically significant difference was detected in the longitudinal change in IOP, the difference between each time interval was compared using the paired t test with the Bonferroni adjustment for multiple comparisons ( P < .0024). A difference in postoperative longitudinal change in IOP between groups was compared with a mixed linear model using incision type, time, and incision-by-time interaction as covariates. Differences in postoperative IOP between groups were also compared using a mixed linear model. Differences with a P value of less than .05 were considered statistically significant.




Results


The 64 enrolled patients completed all scheduled examinations. The procedures of medication, anesthesia, and surgery were almost the same between groups, and therefore the patients were unaware of which eye had undergone scleral tunnel incision or clear corneal incision. The examiners were unaware of which eyes had undergone scleral tunnel incision or clear corneal incision because the ocular and wound appearances were similar when the IOP was measured. Because the clinical research coordinator concealed the assignment schedule until all data were collected, the data analyst did not know which eyes were assigned to the scleral tunnel incision or clear corneal incision groups.


Mean (±standard deviation) patient age was 69.3 ± 5.7 years (range, 48–78 years), and there were 21 men and 43 women. Patient characteristics at baseline and surgical factors of the scleral tunnel incision and clear corneal incision groups are shown in Table 1 . The ratio of left to right eyes, manifest spherical equivalent value, corneal astigmatism, nuclear density, central corneal thickness, uncorrected or distance-corrected visual acuity, mean ultrasound time, mean ultrasound power, mean total ultrasound power emitted, and irrigation volume were not significantly different between groups ( P ≥ .2430). Mean surgical time was significantly longer in the scleral tunnel incision group than in the clear corneal incision group ( P < .0001).



Table 1

Comparison of Patient Characteristics at Baseline and Surgical Factors of the Scleral Tunnel Incision and Clear Corneal Incision Groups










































































STI Group CCI Group P Value
Patient characteristics at baseline
Age (y) 69.3 ± 5.8 69.3 ± 5.8
Keratometric cylinder (D) 0.55 ± 0.41 0.61 ± 0.36 .2299
MRSE (D) −0.89 ± 3.53 −0.90 ± 3.14 .8656
LogMAR visual acuity 0.39 ± 0.18 0.38 ± 0.20 .7160
Preoperative IOP (mm Hg) 15.1 ± 1.7 14.6 ± 1.8 .1632
Nuclear firmness 2.2 ± 0.3 2.1 ± 0.3 .6588
Surgical factors
Surgery time (min) 9.6 ± 2.1 8.2 ± 1.6 <.0001 a
Ultrasound time (s) 32.5 ± 11.5 32.7 ± 10.2 .6625
Ultrasound power (%) 21.5 ± 4.9 22.1 ± 4.6 .4945
Total ultrasound energy b 46.7 ± 6.7 46.3 ± 6.6 .6926
Irrigation volume (mL) 57.2 ± 12.2 55.5 ± 11.3 .6475

CCI = clear corneal incision; D = diopter; IOP = intraocular pressure; LogMAR = logarithm of minimal angle of resolution; MRSE = manifest refraction spherical equivalent; STI = scleral tunnel incision.

a Statistically significant difference.


b Total ultrasound power emitted (millijoules).



Stromal hydration around the wound was not performed in any of the eyes in the scleral tunnel incision group and was performed in 56 of 64 eyes (87.5%) in the clear corneal incision group ( P < .0001; Table 2 ). Mean IOP changed significantly in both groups during the follow-up periods ( P < .0001). In the scleral tunnel incision group, after adjusting the IOP to 21.1 ± 2.7 mm Hg at the conclusion of surgery, mean IOP decreased to the preoperative level within 15 minutes postoperatively and thereafter did not change significantly for up to 60 minutes postoperatively ( Figure 1 ). In the clear corneal incision group, after adjusting the IOP to 20.6 ± 3.3 mm Hg, IOP returned to the preoperative level within 15 minutes postoperatively, and thereafter further reduced to levels significantly lower than the preoperative IOP at 30 minutes postoperatively ( Figure 2 ). The mixed linear model analysis showed that postoperative longitudinal changes in IOP did not significantly differ between the scleral tunnel incision and the clear corneal incision groups ( P = .3229).



Table 2

Number (Percentage) of Eyes With Intraocular Pressure Lower Than 10 mm Hg or 5 mm Hg, and Number (Percentage) of Eyes Requiring Wound Hydration for Intraocular Pressure Adjustment





























STI Group(n = 64) CCI Group(n = 64) P Value
Intraocular pressure
<10 mm Hg 0 7 (10.9%) .0131 a
<5 mm Hg 0 0 >.9999
Wound hydration 0 56 (87.5%) <.0001 a

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jan 8, 2017 | Posted by in OPHTHALMOLOGY | Comments Off on Intraocular Pressure and Wound Status in Eyes Immediately After Scleral Tunnel Incision and Clear Corneal Incision Cataract Surgery

Full access? Get Clinical Tree

Get Clinical Tree app for offline access